Test Catalog

Test Id : JAK2M

JAK2 V617F Mutation Detection, Bone Marrow

Useful For
Suggests clinical disorders or settings where the test may be helpful

Aiding in the distinction between a reactive blood cytosis and a chronic myeloproliferative disorder using bone marrow specimens

Testing Algorithm
Delineates situations when tests are added to the initial order. This includes reflex and additional tests.

Method Name
A short description of the method used to perform the test

Quantitative Polymerase Chain Reaction (PCR)

NY State Available
Indicates the status of NY State approval and if the test is orderable for NY State clients.

Yes

Reporting Name
Lists a shorter or abbreviated version of the Published Name for a test

JAK2 V617F Mutation Detection, BM

Aliases
Lists additional common names for a test, as an aid in searching

Janus kinase 2 gene

Tyrosine Kinase Mutation

Testing Algorithm
Delineates situations when tests are added to the initial order. This includes reflex and additional tests.

Specimen Type
Describes the specimen type validated for testing

Bone Marrow

Shipping Instructions

Specimen must arrive within 7 days of collection.

Specimen Required
Defines the optimal specimen required to perform the test and the preferred volume to complete testing

Container/Tube:

Preferred: Lavender top (EDTA)

Acceptable: Yellow top (ACD)

Specimen Volume: 2 mL

Collection Instructions:

1. Invert several times to mix bone marrow.

2. Send bone marrow specimen in original tube. Do not aliquot.

Special Instructions
Library of PDFs including pertinent information and forms related to the test

Forms

1. Hematopathology Patient Information (T676)

2. If not ordering electronically, complete, print, and send a Hematopathology/Cytogenetics Test Request (T726) with the specimen.

Specimen Minimum Volume
Defines the amount of sample necessary to provide a clinically relevant result as determined by the Testing Laboratory

1 mL

Reject Due To
Identifies specimen types and conditions that may cause the specimen to be rejected

Gross hemolysis Reject
Moderately to severely clotted Reject

Specimen Stability Information
Provides a description of the temperatures required to transport a specimen to the performing laboratory, alternate acceptable temperatures are also included

Specimen Type Temperature Time Special Container
Bone Marrow Ambient (preferred) 7 days PURPLE OR PINK TOP/EDTA
Refrigerated 7 days PURPLE OR PINK TOP/EDTA

Useful For
Suggests clinical disorders or settings where the test may be helpful

Aiding in the distinction between a reactive blood cytosis and a chronic myeloproliferative disorder using bone marrow specimens

Testing Algorithm
Delineates situations when tests are added to the initial order. This includes reflex and additional tests.

Clinical Information
Discusses physiology, pathophysiology, and general clinical aspects, as they relate to a laboratory test

The Janus kinase 2 (JAK2) gene codes for a tyrosine kinase (JAK2) associated with the cytoplasmic portion of a variety of transmembrane cytokine and growth factor receptors important for signal transduction in hematopoietic cells. Signaling via JAK2 activation causes phosphorylation of downstream signal transducers and activators of transcription (STAT) proteins (eg, STAT5) ultimately leading to cell growth and differentiation. BCR-ABL1-negative myeloproliferative neoplasms (MPN) frequently harbor an acquired single nucleotide variant in JAK2 characterized as c.G1849T; p. Val617Phe (V617F). This variant is identified overall in approximately two-thirds of all MPN,(1-3) but the prevalence varies by MPN subtype. The JAK2 V617F variant is present in 95% to 98% of polycythemia vera patients, 50% to 60% of primary myelofibrosis (PMF) patients, and 50% to 60% of essential thrombocythemia (ET) patients. It has infrequently been described in other myeloid neoplasms, including chronic myelomonocytic leukemia and myelodysplastic syndrome.(4) This variant is not seen in chronic myelogenous leukemia or reactive conditions with elevated blood counts. Detection of the JAK2 V617F variant is useful to help establish the diagnosis of MPN. However, a negative JAK2 V617F result does not indicate absence of an MPN. Other important molecular markers in BCR-ABL1-negative MPN include CALR exon 9 variant (20%-30% of PMF and ET) and MPL exon 10 variant (5%-10% of PMF and 3%-5% of ET). Variants in JAK2, CALR, and MPL are essentially mutually exclusive.

Reference Values
Describes reference intervals and additional information for interpretation of test results. May include intervals based on age and sex when appropriate. Intervals are Mayo-derived, unless otherwise designated. If an interpretive report is provided, the reference value field will state this.

An interpretive report will be provided.

Interpretation
Provides information to assist in interpretation of the test results

The results will be reported as 1 of the 2 states:

-Negative for JAK2 V617F variant

-Positive for JAK2 V617F variant

 

Positive variant status is highly suggestive of a myeloid neoplasm but must be correlated with clinical and other laboratory features for definitive diagnosis.

 

Negative variant status does not exclude the presence of a myeloproliferative neoplasm or other neoplasm.

 

Results below the laboratory cutoff for positivity are of unclear clinical significance at this time.

Cautions
Discusses conditions that may cause diagnostic confusion, including improper specimen collection and handling, inappropriate test selection, and interfering substances

A positive result is not specific for a particular subtype of myeloproliferative neoplasm and clinicopathologic correlation is necessary in all cases. If this test is ordered in the setting of erythrocytosis and suspicion of polycythemia vera, interpretation requires correlation with a concurrent or recent prior bone marrow evaluation.

 

A negative result does not exclude the presence of a myeloproliferative neoplasm or other neoplastic process.

 

In rare cases, a variant other than the V617F may be present in an area that interferes with primer or probe binding and cause a false-negative result.

Supportive Data

Analytical sensitivity is determined at 0.06% (by dilution of a JAK2 V617F-positive cell line DNA into a negative cell line DNA).

Clinical Reference
Recommendations for in-depth reading of a clinical nature

1. Baxter EJ, Scott LM, Campbell PJ, et al: Acquired mutation of the tyrosine kinase JAK2 in human myeloproliferative disorders. Lancet. 2005 Mar;365(9464):1054-1061

2. James C, Ugo V, Le Couedic JP, et al: A unique clonal JAK2 mutation leading to constitutive signaling causes polycythaemia vera. Nature. 2005 Apr 28;434(7037):1144-1148

3. Kralovics R, Passamonti F, Buser AS, et al: A gain-of-function mutation of JAK2 in myeloproliferative disorders. N Engl J Med. 2005 Apr 28;352(17):1779-1790

4. Steensma DP, Dewald GW, Lasho TL, et al: The JAK2 V617F activating tyrosine kinase mutation is an infrequent event in both "atypical" myeloproliferative disorders and the myelodysplastic syndromes. Blood. 2005 Aug 15;106(4):1207-1209

5. Gong JZ, Cook JR, Greiner TC, et al: Laboratory practice guidelines for detecting and reporting JAK2 and MPL mutations in myeloproliferative neoplasms: a report of the Association for Molecular Pathology. J Mol Diagn. 2013 Nov;15(6):733-744

Method Description
Describes how the test is performed and provides a method-specific reference

Genomic DNA is extracted, and 2 polymerase chain reactions (PCR) are used for each sample. In each reaction, a short fragment of genomic DNA, including the variant site, is amplified using quantitative PCR in a real-time PCR instrument. In the first reaction, the 5' terminal base of the reverse primer matches the mutated sequence, and the PCR conditions are such that it will only bind mutated DNA. In the second reaction, the 5' terminal base of the reverse primer matches the wildtype sequence, and the PCR conditions are such that it will only bind the wild-type sequence. In both reactions, the PCR is monitored using TaqMan probe chemistry. The amount of mutated DNA and the amount of wildtype DNA is measured for each sample. In each run, the amount of mutated and wild-type DNA in a calibrator DNA sample is also measured. The calibrator is a mixture of DNA from a positive cell line (HEL) and a negative cell line (HL60) that is frozen in aliquots and expected to give an identical result in each run. Deviations in the calibrator result are assumed to be due to deviations in the run conditions and the sample results are corrected accordingly. Following each reaction, Relative Quantification Software is used to calculate the normalized mutated:wildtype ratio, which is expressed as a unitless ratio following correction with the calibrator data.

 

The formula for the normalized ratio is as follows:

 

Normalized ratio =

mutated/wildtype (sample)

mutated/wildtype (calibrator)

 

The final result is reported as percent JAK2 V617F of total JAK2 (ie, [mutated/mutated + wildtype] x 100%), calculated from the normalized mutated:wildtype ratio.(Unpublished Mayo method)

PDF Report
Indicates whether the report includes an additional document with charts, images or other enriched information

No

Day(s) Performed
Outlines the days the test is performed. This field reflects the day that the sample must be in the testing laboratory to begin the testing process and includes any specimen preparation and processing time before the test is performed. Some tests are listed as continuously performed, which means that assays are performed multiple times during the day.

Monday through Friday

Report Available
The interval of time (receipt of sample at Mayo Clinic Laboratories to results available) taking into account standard setup days and weekends. The first day is the time that it typically takes for a result to be available. The last day is the time it might take, accounting for any necessary repeated testing.

2 to 5 days

Specimen Retention Time
Outlines the length of time after testing that a specimen is kept in the laboratory before it is discarded

Bone marrow: 2 weeks; Extracted DNA: 3 months

Performing Laboratory Location
Indicates the location of the laboratory that performs the test

Rochester

Fees
Several factors determine the fee charged to perform a test. Contact your U.S. or International Regional Manager for information about establishing a fee schedule or to learn more about resources to optimize test selection.

  • Authorized users can sign in to Test Prices for detailed fee information.
  • Clients without access to Test Prices can contact Customer Service 24 hours a day, seven days a week.
  • Prospective clients should contact their account representative. For assistance, contact Customer Service.

Test Classification
Provides information regarding the medical device classification for laboratory test kits and reagents. Tests may be classified as cleared or approved by the US Food and Drug Administration (FDA) and used per manufacturer instructions, or as products that do not undergo full FDA review and approval, and are then labeled as an Analyte Specific Reagent (ASR) product.

This test was developed and its performance characteristics determined by Mayo Clinic in a manner consistent with CLIA requirements. It has not been cleared or approved by the US Food and Drug Administration.

CPT Code Information
Provides guidance in determining the appropriate Current Procedural Terminology (CPT) code(s) information for each test or profile. The listed CPT codes reflect Mayo Clinic Laboratories interpretation of CPT coding requirements. It is the responsibility of each laboratory to determine correct CPT codes to use for billing.

CPT codes are provided by the performing laboratory.

81270-JAK2 (Janus kinase 2) (eg, myeloproliferative disorder) gene analysis, p.Val617Phe (V617F) variant

LOINC® Information
Provides guidance in determining the Logical Observation Identifiers Names and Codes (LOINC) values for the order and results codes of this test. LOINC values are provided by the performing laboratory.

Test Id Test Order Name Order LOINC Value
JAK2M JAK2 V617F Mutation Detection, BM 72333-8
Result Id Test Result Name Result LOINC Value
Applies only to results expressed in units of measure originally reported by the performing laboratory. These values do not apply to results that are converted to other units of measure.
31157 JAK2 V617F Mutation Detection, BM 72333-8
39723 JAK2 Result 53761-3

Test Setup Resources

Setup Files
Test setup information contains test file definition details to support order and result interfacing between Mayo Clinic Laboratories and your Laboratory Information System.

Excel | PHP Pdf | CMS Pdf

Sample Reports
Normal and Abnormal sample reports are provided as references for report appearance.

Normal Reports | Abnormal Reports

SI Sample Reports
International System (SI) of Unit reports are provided for a limited number of tests. These reports are intended for international account use and are only available through MayoLINK accounts that have been defined to receive them.

SI Normal Reports | SI Abnormal Reports