Test Catalog

Test Id : CRSY

Chromium, Synovial Fluid

Useful For
Suggests clinical disorders or settings where the test may be helpful

Monitoring metallic prosthetic implant wear and local tissue destruction in failed hip arthroplasty constructs

 

This test is not useful for assessment of potential chromium toxicity.

Method Name
A short description of the method used to perform the test

Inductively Coupled Plasma Mass Spectrometry (ICP-MS)

NY State Available
Indicates the status of NY State approval and if the test is orderable for NY State clients.

Yes

Reporting Name
Lists a shorter or abbreviated version of the Published Name for a test

Chromium, Synovial Fl

Aliases
Lists additional common names for a test, as an aid in searching

Chromium (Cr)

Cr (Chromium)

Specimen Type
Describes the specimen type validated for testing

Synovial Fluid

Specimen Required
Defines the optimal specimen required to perform the test and the preferred volume to complete testing

Patient Preparation: High concentrations of gadolinium and iodine are known to interfere with most inductively coupled plasma mass spectrometry-based metal tests. If either gadolinium- or iodine-containing contrast media has been administered, a specimen should not be collected for at least 96 hours.

Supplies: Metal Free B-D Tube (EDTA), 6 mL (T183)

Container/Tube: Royal blue top (metal-free EDTA)

Specimen Volume: 1 mL

Collection Instructions: See Metals Analysis Specimen Collection and Transport for complete instructions.

Additional Information: Chromium is present in the black rubber plunger seals found in most disposable syringes. As a result, synovial fluid should not be collected in these devices as contamination may occur.

Special Instructions
Library of PDFs including pertinent information and forms related to the test

Specimen Minimum Volume
Defines the amount of sample necessary to provide a clinically relevant result as determined by the Testing Laboratory

0.4 mL

Reject Due To
Identifies specimen types and conditions that may cause the specimen to be rejected

Gross hemolysis OK
Gross lipemia OK
Gross icterus OK

Specimen Stability Information
Provides a description of the temperatures required to transport a specimen to the performing laboratory, alternate acceptable temperatures are also included

Specimen Type Temperature Time Special Container
Synovial Fluid Refrigerated (preferred) 90 days
Ambient 90 days
Frozen 90 days

Useful For
Suggests clinical disorders or settings where the test may be helpful

Monitoring metallic prosthetic implant wear and local tissue destruction in failed hip arthroplasty constructs

 

This test is not useful for assessment of potential chromium toxicity.

Clinical Information
Discusses physiology, pathophysiology, and general clinical aspects, as they relate to a laboratory test

Chromium (Cr) is a naturally-occurring element widely distributed in the environment. It exists in several valence states with the 3 main forms being Cr(0), Cr(III), and Cr(VI). Cr(III) is an essential trace element that enhances the action of insulin. Deficiency leads to impaired growth, reduced life span, corneal lesions, and alterations in carbohydrates, lipid, and protein metabolism.

 

Chromium is widely used in manufacturing processes to make various metal alloys, such as stainless steel. It is also used in many consumer products, including wood treated with copper dichromate, leather tanned with chromic sulfate, and metal-on-metal hip replacements.

 

Per US Food and Drug Administration recommendations, orthopedic surgeons should consider measuring and following serial chromium concentrations in EDTA anticoagulated whole blood in symptomatic patients with metal-on-metal hip implants as part of their overall clinical evaluation. However, a recent publication(1) has shown synovial fluid measurements were superior to whole blood and serum chromium concentrations in predicting local tissue destruction in failed hip arthroplasty constructs.

 

Prosthetic devices produced by Depuy Company, Dow Corning, Howmedica, LCS, PCA, Osteonics, Richards Company, Tricon, and Whiteside are typically made of chromium, cobalt, and molybdenum. This list of products is incomplete, and the products' compositions change occasionally; see each prostheses' product information for composition details.

 

The general population is most likely to be exposed to trace levels of chromium, as Cr(III) is naturally occurring in foods, such as fruits, vegetables, nuts, beverages, and meats. The highest potential occupational exposure occurs in the metallurgy and tanning industries, where workers may be exposed to high air concentrations. In cases of suspected toxicity, blood, serum, or urine concentrations of chromium can be checked.

Reference Values
Describes reference intervals and additional information for interpretation of test results. May include intervals based on age and sex when appropriate. Intervals are Mayo-derived, unless otherwise designated. If an interpretive report is provided, the reference value field will state this.

0-17 years: Not established

> or =18 years: <16.9 ng/mL

Interpretation
Provides information to assist in interpretation of the test results

Based on an internal study, synovial fluid chromium concentrations of 16.9 ng/mL or above were more likely due to a metal reaction (eg, adverse local tissue reaction [ALTR]/adverse reaction to metal debris [ARMD]) versus a nonmetal reaction in patients undergoing metal-on-metal revision (sensitivity of 92.3% and specificity of 92.6%).

Cautions
Discusses conditions that may cause diagnostic confusion, including improper specimen collection and handling, inappropriate test selection, and interfering substances

This test is intended for monitoring of metallic prosthetic implant wear and should not be ordered to assess potential chromium toxicity.

 

Chromium is present in the environment at 100-fold to 1000-fold higher concentration than found in biological tissues. Reports of increased chromium could be due to external contamination. Metal-free blood collection procedures must be followed. Elevated trace element concentrations in the absence of corroborating clinical information do not independently predict prosthesis wear or failure.

 

This test should only be used in individuals with metallic prosthetic implants. The significance of chromium concentrations in synovial fluid in patients without implants is unknown.

Clinical Reference
Recommendations for in-depth reading of a clinical nature

1. Houdek MT, Taunton MJ, Wyles CC, Jannetto PJ, Lewallen DG, Berry DJ. Synovial fluid metal ion levels are superior to blood metal ion levels in predicting an adverse local tissue reaction in failed total hip arthroplasty. J Arthroplasty. 2021;36(9):3312-3317.e1. doi:10.1016/j.arth.2021.04.034

2. Eltit F, Assiri A, Garbuz D, et al. Adverse reactions to metal on polyethylene implants: Highly destructive lesions related to elevated concentration of cobalt and chromium in synovial fluid. J Biomed Mater Res A. 2017;105(7):1876-1886. doi:10.1002/jbm.a.36057

3. Lass R, Grubl A, Kolb A, et al. Comparison of synovial fluid, urine, and serum ion levels in metal-on-metal total hip arthroplasty at minimum follow-up of 18 years. J Orthop Res. 2014;32(9):1234-1240. doi:10.1002/jor.22652

4. De Pasquale D, Stea S, Squarzoni S, et al. Metal-on-metal hip prostheses: Correlation between debris in the synovial fluid and levels of cobalt and chromium ions in the bloodstream. Int Orthop. 2014;38(3):469-475. doi:10.1007/s00264-013-2137-5

Method Description
Describes how the test is performed and provides a method-specific reference

The metal of interest is analyzed by inductively coupled plasma mass spectrometry.(Unpublished Mayo method)

PDF Report
Indicates whether the report includes an additional document with charts, images or other enriched information

No

Day(s) Performed
Outlines the days the test is performed. This field reflects the day that the sample must be in the testing laboratory to begin the testing process and includes any specimen preparation and processing time before the test is performed. Some tests are listed as continuously performed, which means that assays are performed multiple times during the day.

Friday

Report Available
The interval of time (receipt of sample at Mayo Clinic Laboratories to results available) taking into account standard setup days and weekends. The first day is the time that it typically takes for a result to be available. The last day is the time it might take, accounting for any necessary repeated testing.

2 to 8 days

Specimen Retention Time
Outlines the length of time after testing that a specimen is kept in the laboratory before it is discarded

14 days

Performing Laboratory Location
Indicates the location of the laboratory that performs the test

Rochester

Fees
Several factors determine the fee charged to perform a test. Contact your U.S. or International Regional Manager for information about establishing a fee schedule or to learn more about resources to optimize test selection.

  • Authorized users can sign in to Test Prices for detailed fee information.
  • Clients without access to Test Prices can contact Customer Service 24 hours a day, seven days a week.
  • Prospective clients should contact their account representative. For assistance, contact Customer Service.

Test Classification
Provides information regarding the medical device classification for laboratory test kits and reagents. Tests may be classified as cleared or approved by the US Food and Drug Administration (FDA) and used per manufacturer instructions, or as products that do not undergo full FDA review and approval, and are then labeled as an Analyte Specific Reagent (ASR) product.

This test was developed and its performance characteristics determined by Mayo Clinic in a manner consistent with CLIA requirements. It has not been cleared or approved by the US Food and Drug Administration.

CPT Code Information
Provides guidance in determining the appropriate Current Procedural Terminology (CPT) code(s) information for each test or profile. The listed CPT codes reflect Mayo Clinic Laboratories interpretation of CPT coding requirements. It is the responsibility of each laboratory to determine correct CPT codes to use for billing.

CPT codes are provided by the performing laboratory.

82495

LOINC® Information
Provides guidance in determining the Logical Observation Identifiers Names and Codes (LOINC) values for the order and results codes of this test. LOINC values are provided by the performing laboratory.

Test Id Test Order Name Order LOINC Value
CRSY Chromium, Synovial Fl 95526-0
Result Id Test Result Name Result LOINC Value
Applies only to results expressed in units of measure originally reported by the performing laboratory. These values do not apply to results that are converted to other units of measure.
606353 Chromium, Synovial Fl 95526-0

Test Setup Resources

Setup Files
Test setup information contains test file definition details to support order and result interfacing between Mayo Clinic Laboratories and your Laboratory Information System.

Excel | PHP Pdf | CMS Pdf

Sample Reports
Normal and Abnormal sample reports are provided as references for report appearance.

Normal Reports | Abnormal Reports

SI Sample Reports
International System (SI) of Unit reports are provided for a limited number of tests. These reports are intended for international account use and are only available through MayoLINK accounts that have been defined to receive them.

SI Normal Reports | SI Abnormal Reports