Test Id : CA153

Cancer Antigen 15-3 (CA 15-3), Serum

Useful For
Suggests clinical disorders or settings where the test may be helpful

Managing breast cancer patients when used in conjunction with clinical information and other diagnostic procedures

 

Serial testing to assist in early detection of disease recurrence in previously treated stage II and III breast cancer patients

 

Monitoring response to therapy in metastatic breast cancer patients

 

This test is not useful as a cancer screening test.

Method Name
A short description of the method used to perform the test

Electrochemiluminescence Immunoassay (ECLIA)

NY State Available
Indicates the status of NY State approval and if the test is orderable for NY State clients.

Yes

Reporting Name
Lists a shorter or abbreviated version of the Published Name for a test

Cancer Ag 15-3, (CA 15-3), S

Aliases
Lists additional common names for a test, as an aid in searching

CA 15-3 (Cancer Antigen 15-3)

CA153

MUC-1

Mucin-Like Carcinoma-Associated Antigen

Specimen Type
Describes the specimen type validated for testing

Serum

Specimen Required
Defines the optimal specimen required to perform the test and the preferred volume to complete testing

Patient Preparation: For 12 hours before specimen collection do not take multivitamins or dietary supplements containing biotin (vitamin B7), which is commonly found in hair, skin, and nail supplements and multivitamins.

Supplies: Sarstedt 5 mL Aliquot Tube (T914)

Collection Container/Tube:

Preferred: Serum gel

Acceptable: Red top

Submission Container/Tube: Plastic vial

Specimen Volume: 1 mL

Collection Instructions: Centrifuge and aliquot serum into plastic vial.

Forms

Specimen Minimum Volume
Defines the amount of sample necessary to provide a clinically relevant result as determined by the Testing Laboratory

0.75 mL

Reject Due To
Identifies specimen types and conditions that may cause the specimen to be rejected

Gross hemolysis Reject
Gross lipemia OK

Specimen Stability Information
Provides a description of the temperatures required to transport a specimen to the performing laboratory, alternate acceptable temperatures are also included

Specimen Type Temperature Time Special Container
Serum Refrigerated (preferred) 7 days
Frozen 90 days

Useful For
Suggests clinical disorders or settings where the test may be helpful

Managing breast cancer patients when used in conjunction with clinical information and other diagnostic procedures

 

Serial testing to assist in early detection of disease recurrence in previously treated stage II and III breast cancer patients

 

Monitoring response to therapy in metastatic breast cancer patients

 

This test is not useful as a cancer screening test.

Clinical Information
Discusses physiology, pathophysiology, and general clinical aspects, as they relate to a laboratory test

Carcinoma of the breast is the most prevalent form of cancer in women. These tumors often produce mucinous antigens, which are large molecular weight glycoproteins with O-linked oligosaccharide chains. Tumor-associated antigens encoded by the human MUC-1 gene are known by several names, including MAM6, milk mucin antigen, cancer antigen (CA) 27.29, and CA 15-3.

 

CA 15-3 assay values are not elevated in most normal individuals and are frequently elevated in sera from breast cancer patients.

 

Nonmammary malignancies in which elevated CA 15-3 assay values have been reported include: lung, colon, pancreas, primary liver, ovary, cervix, and endometrium.

Reference Values
Describes reference intervals and additional information for interpretation of test results. May include intervals based on age and sex when appropriate. Intervals are Mayo-derived, unless otherwise designated. If an interpretive report is provided, the reference value field will state this.

Males: <30 U/mL (use not defined)

Females: <30 U/mL

Interpretation
Provides information to assist in interpretation of the test results

Increasing and decreasing values show correlation with disease progression and regression, respectively.(1) Increasing cancer antigen 15-3 (CA 15-3) assay values in patients at risk for breast cancer recurrence after primary therapy may be indicative of recurrent disease before it can be detected clinically (2,3) and may be used as an indication that additional tests or procedures should be performed.

Cautions
Discusses conditions that may cause diagnostic confusion, including improper specimen collection and handling, inappropriate test selection, and interfering substances

Testing for cancer antigen 15-3 (CA 15-3) should be performed in conjunction with other clinical methods used for the early detection of recurrence.

 

Some patients who have been exposed to mouse antigens, whether in the environment or as part of treatment or imaging procedures, may have circulating antimouse antibodies. These antibodies may interfere with the assay reagents to produce unreliable CA 15-3 assay results.

 

In rare cases, interference due to extremely high titers of antibodies to ruthenium or streptavidin can occur.

Clinical Reference
Recommendations for in-depth reading of a clinical nature

1. Molina R, Zanon G, Filella X, et al: Use of serial carcinoembryonic antigen and CA 15-3 assays in detecting relapses in breast cancer patients. Breast Cancer Res Treat. 1995 Jan;36:41-48

2. Geraghty JG, Coveney EC, Sherry F, O'Higgins NJ, Duffy MJ: CA 15-3 in patients with locoregional and metastatic breast carcinoma. Cancer. 1992 Dec 15;70(12):2831-2834

3. Kallioniemi OP, Oksa H, Aaran RK, Hietanen T, Lehtinen M, Koivula T: Serum CA 15-3 assay in the diagnosis and follow-up of breast cancer. Br J Cancer. 1988 Aug;58(2):213-215

4. Lin DC, Genzen JR: Concordance analysis of paired cancer antigen (CA) 15-3 and 27.29 testing. Breast Cancer Research and Treatment. 2018;167:269-276

Method Description
Describes how the test is performed and provides a method-specific reference

The Roche cancer antigen 15-3 (CA 15-3) method is a sandwich electrochemiluminescence immunoassay that employs a biotinylated monoclonal CA 15-3-specific antibody and a monoclonal CA 15-3-specific antibody. CA 15-3 in the automatically prediluted specimen reacts with both the biotinylated monoclonal CA 15-3-specific antibody (mouse) and the monoclonal CA 15-3-specific antibody (mouse) labeled with a ruthenium complex, forming a sandwich complex. Streptavidin-coated microparticles are added and the mixture is aspirated into the measuring cell where the microparticles are magnetically captured onto the surface of the electrode. Unbound substances are then removed with ProCell. Application of voltage to the electrode induces the chemiluminescent emission, which is then measured.(Package insert: Roche CA 15-3 reagent. Roche Diagnostics; V14 09/2010)

PDF Report
Indicates whether the report includes an additional document with charts, images or other enriched information

No

Day(s) Performed
Outlines the days the test is performed. This field reflects the day that the sample must be in the testing laboratory to begin the testing process and includes any specimen preparation and processing time before the test is performed. Some tests are listed as continuously performed, which means that assays are performed multiple times during the day.

Monday through Friday

Report Available
The interval of time (receipt of sample at Mayo Clinic Laboratories to results available) taking into account standard setup days and weekends. The first day is the time that it typically takes for a result to be available. The last day is the time it might take, accounting for any necessary repeated testing.

1 to 3 days

Specimen Retention Time
Outlines the length of time after testing that a specimen is kept in the laboratory before it is discarded

3 months

Performing Laboratory Location
Indicates the location of the laboratory that performs the test

Rochester

Fees
Several factors determine the fee charged to perform a test. Contact your U.S. or International Regional Manager for information about establishing a fee schedule or to learn more about resources to optimize test selection.

  • Authorized users can sign in to Test Prices for detailed fee information.
  • Clients without access to Test Prices can contact Customer Service 24 hours a day, seven days a week.
  • Prospective clients should contact their account representative. For assistance, contact Customer Service.

Test Classification
Provides information regarding the medical device classification for laboratory test kits and reagents. Tests may be classified as cleared or approved by the US Food and Drug Administration (FDA) and used per manufacturer instructions, or as products that do not undergo full FDA review and approval, and are then labeled as an Analyte Specific Reagent (ASR) product.

This test has been cleared, approved, or is exempt by the US Food and Drug Administration and is used per manufacturer's instructions. Performance characteristics were verified by Mayo Clinic in a manner consistent with CLIA requirements.

CPT Code Information
Provides guidance in determining the appropriate Current Procedural Terminology (CPT) code(s) information for each test or profile. The listed CPT codes reflect Mayo Clinic Laboratories interpretation of CPT coding requirements. It is the responsibility of each laboratory to determine correct CPT codes to use for billing.

CPT codes are provided by the performing laboratory.

86300

LOINC® Information
Provides guidance in determining the Logical Observation Identifiers Names and Codes (LOINC) values for the order and results codes of this test. LOINC values are provided by the performing laboratory.

Test Id Test Order Name Order LOINC Value
CA153 Cancer Ag 15-3, (CA 15-3), S 83083-6
Result Id Test Result Name Result LOINC Value
Applies only to results expressed in units of measure originally reported by the performing laboratory. These values do not apply to results that are converted to other units of measure.
CA153 Cancer Ag 15-3, (CA 15-3), S 83083-6

Test Setup Resources

Setup Files
Test setup information contains test file definition details to support order and result interfacing between Mayo Clinic Laboratories and your Laboratory Information System.

Excel | PHP Pdf | CMS Pdf

Sample Reports
Normal and Abnormal sample reports are provided as references for report appearance.

Normal Reports | Abnormal Reports

SI Sample Reports
International System (SI) of Unit reports are provided for a limited number of tests. These reports are intended for international account use and are only available through MayoLINK accounts that have been defined to receive them.

SI Normal Reports | SI Abnormal Reports