Test Catalog

Test Id : HERMB

HER2, Breast, Semi-Quantitative Immunohistochemistry, Manual with HER2 FISH Reflex

Useful For
Suggests clinical disorders or settings where the test may be helpful

Determining overexpression of HER2 protein on formalin-fixed, paraffin-embedded tissue sections

Reflex Tests
Lists tests that may or may not be performed, at an additional charge, depending on the result and interpretation of the initial tests.

Test Id Reporting Name Available Separately Always Performed
HERBN HER Breast IHC Automated NO Reflex Yes No
H2BR HER2, Breast Tumor, FISH, Tissue Yes No

Testing Algorithm
Delineates situations when tests are added to the initial order. This includes reflex and additional tests.

Cases that are equivocal (2+) by immunohistochemical stain will reflex to HER2 amplification by fluorescence in situ hybridization at an additional charge.

Method Name
A short description of the method used to perform the test

Immunoperoxidase Stain with Manual Quantitative Immunohistochemistry (IHC)

NY State Available
Indicates the status of NY State approval and if the test is orderable for NY State clients.

Yes

Reporting Name
Lists a shorter or abbreviated version of the Published Name for a test

HER Breast SemiQuant IHC + Reflex

Aliases
Lists additional common names for a test, as an aid in searching

c-erbB2 Immunoperoxidase Stain

c-neu Immunoperoxidase Stain

Her-2

HER2-Neu

HER2/Neu

HER2Neu

Testing Algorithm
Delineates situations when tests are added to the initial order. This includes reflex and additional tests.

Cases that are equivocal (2+) by immunohistochemical stain will reflex to HER2 amplification by fluorescence in situ hybridization at an additional charge.

Specimen Type
Describes the specimen type validated for testing

Special

Ordering Guidance

For gastroesophageal cancer, order HERGM / HER2, Gastric/Esophageal, Semi-Quantitative Immunohistochemistry, Manual or HERGN / HER2, Gastric/Esophageal, Semi-Quantitative Immunohistochemistry, Manual, No Reflex.

Shipping Instructions

Attach the green "Attention Pathology" address label (T498) to the outside of the transport container before putting into the courier mailer.

Necessary Information

1. An accompanying pathology report stating the final diagnosis is required. If not available, a preliminary diagnosis is acceptable only if it refers to invasive or metastatic breast carcinoma.

2. Information regarding fixative used, time to fixation, and duration of fixation as well as tumor type and classification is required.

The following questions, as stated on the Pathology/Cytology Information (T707) form or presented electronically, must be answered:

 a. "Was specimen fixed in 10% neutral buffered formalin within 1 hour from surgical collection time? Yes, No, or Unknown."

 b. "Has specimen been fixed in 10% neutral buffered formalin for 6 to 72 hours? Yes, No, or Unknown."

 c. "Was tissue decalcified? Yes, No or Unknown."

 d. "Tumor type? Primary invasive breast carcinoma or metastatic breast carcinoma."

 e. "Tumor classification? Invasive breast carcinoma, metastatic breast carcinoma, or micro-invasive breast carcinoma."

ORDER QUESTIONS AND ANSWERS

Question ID Description Answers
MA058 Fixed in 10% NB formalin w/in 1 hr Yes
No
Unknown
MA059 Fixed in 10% NB formalin 6-72 hrs
MA060 Tumor type Primary invasive breast carcinoma
Metastatic breast carcinoma
MA061 Tumor classification Invasive breast carcinoma
Metastatic breast carcinoma
Micro-invasive breast carcinoma
MA062 Tissue was decalcified

Specimen Required
Defines the optimal specimen required to perform the test and the preferred volume to complete testing

Supplies: Pathology Packaging Kit (T554)

Specimen Type:

Preferred: A paraffin-embedded tissue block containing breast cancer tissue that has been fixed in 10% neutral buffered formalin within 1 hour from surgical collection time and for a total of 6 to 72 hours.

Acceptable: 5 Unstained sections containing breast carcinoma on charged slides cut at 4 microns less than 1 month ago. Tissue on the slides should have been fixed in 10% neutral buffered formalin within 1 hour from surgical collection time and for a total of 6 to 72 hours.

Container/Tube: Pathology Packaging Kit

Collection Instructions: Submit paraffin-embedded invasive or metastatic breast carcinoma tissue.

Additional Information:

1. According to the American Society of Clinical Oncology (ASCO)/College of American Pathologists (CAP) guidelines, HER2 protein immunohistochemical test results are only valid for nondecalcified, paraffin-embedded specimens fixed in 10% neutral buffered formalin within 1 hour from surgical collection time and for a total time of 6 to 72 hours. Delay to fixation, or under- or over-fixation may affect these results.(1)

2. HER2 immunohistochemistry testing on intracystic papillary carcinoma and solid papillary carcinoma, without clearly stating invasive carcinoma, is not appropriate and will be reported as indeterminate.

3. Paraffin blocks will be returned with final report.

Special Instructions
Library of PDFs including pertinent information and forms related to the test

Forms

If not ordering electronically, complete, print, and send with the specimen:

1. Pathology/Cytology Information (T707) is required if not ordering electronically.

2. Immunohistochemical (IHC)/In Situ Hybridization (ISH) Stains Request (T763)

Reject Due To
Identifies specimen types and conditions that may cause the specimen to be rejected

No specimen should be rejected.

Specimen Stability Information
Provides a description of the temperatures required to transport a specimen to the performing laboratory, alternate acceptable temperatures are also included

Specimen Type Temperature Time Special Container
Special Ambient (preferred)
Refrigerated

Useful For
Suggests clinical disorders or settings where the test may be helpful

Determining overexpression of HER2 protein on formalin-fixed, paraffin-embedded tissue sections

Testing Algorithm
Delineates situations when tests are added to the initial order. This includes reflex and additional tests.

Cases that are equivocal (2+) by immunohistochemical stain will reflex to HER2 amplification by fluorescence in situ hybridization at an additional charge.

Clinical Information
Discusses physiology, pathophysiology, and general clinical aspects, as they relate to a laboratory test

The HER2 (official gene name ERBB2) proto-oncogene encodes a membrane receptor with tyrosine kinase activity and homology to the epidermal growth factor receptor. Amplification and overexpression of the HER2 gene in human breast, endometrial, ovarian, and other epithelial cancers have been associated with a shorter disease-free interval and shorter overall survival. Overexpression of HER2 protein is an indication for Herceptin (trastuzumab) therapy in patients with breast cancer.

 

This test is most frequently used to evaluate HER2 overexpression in breast cancer.

Reference Values
Describes reference intervals and additional information for interpretation of test results. May include intervals based on age and sex when appropriate. Intervals are Mayo-derived, unless otherwise designated. If an interpretive report is provided, the reference value field will state this.

Reported as negative (0, 1+), equivocal (2+), and strongly positive (3+) according to the interpretation guidelines for the US Food and Drug Administration-approved Ventana Pathway HER2 (4B5) antibody.

Interpretation
Provides information to assist in interpretation of the test results

Results are reported as negative (0, 1+), equivocal (2+), and strongly positive (3+) according to the published American Society of Clinical Oncology (ASCO)/College of American Pathologists (CAP) interpretation guidelines.(1)

Cautions
Discusses conditions that may cause diagnostic confusion, including improper specimen collection and handling, inappropriate test selection, and interfering substances

The performance and quality of immunohistochemical stains in formalin-fixed, paraffin-embedded tissue depends critically on proper fixation.

Clinical Reference
Recommendations for in-depth reading of a clinical nature

1. Wolff AC, Hammond ME, Hicks DG, et al. Human Epidermal Growth Factor Receptor 2 Testing in Breast Cancer: American Society of Clinical Oncology/College of American Pathologists Clinical Practice Guideline Focused Update. J Clin Oncol. 2018;36(20):2105-2122 doi:10.1200/JCO.2018.77.8738

2. Riber-Hansen R, Vainer B, Steiniche T. Digital image analysis: a review of reproducibility, stability and basic requirements for optimal results. Apmis. 2012;120(4):276-289

3. Gavrielides MA, Gallas BD, Lenz P, Badano A, Hewitt SM.: Observer variability in the interpretation of HER2/neu immunohistochemical expression with unaided and computer-aided digital microscopy. Arch Pathol Lab Med. 2011;135(2):233-242

4. Cuadros M, Villegas R. Systematic review of HER2 breast cancer testing. Appl Immunohistochem Mol Morphol. 2009;17(1):1-7

5. Nassar A, Cohen C, Agersborg SS, et al. Trainable immunohistochemical HER2/neu image analysis: a multisite performance study using 260 breast tissue specimens. Arch Pathol Lab Med. 2011;135(7):896-902

Method Description
Describes how the test is performed and provides a method-specific reference

Testing is performed using US Food and Drug Administration-approved Ventana Pathway HER2 (4B5) rabbit monoclonal primary antibody and a proprietary detection system, using modified manufacturer’s instructions.(Package insert: PATHWAY anti-HER-2/neu (4B5) Rabbit Monoclonal Primary Antibody. Ventana Medical Systems, Inc; 09/2022)

 

Scoring is performed according to American Society of Clinical Oncology (ASCO)/College of American Pathologists (CAP) guidelines as follows:

Score of 3+ is defined as circumferential membrane staining that is complete, intense and in greater than 10% of invasive tumor cells.

Score of 2+ is defined as weak to moderate complete membrane staining observed and in greater than 10% of the invasive tumor cells; or circumferential membrane staining that is complete, intense and in less than or equal to 10% of invasive tumor cells.

Score of 1+ is defined as incomplete membrane staining that is faint or barely perceptible and in greater than 10% of the invasive tumor cells; or weak to moderate complete membrane staining observed and less than 10% of the invasive tumor cells.

Score of 0 is defined as no staining observed or membrane staining that is incomplete and is faint or barely perceptible and in less than or equal to 10% of the invasive tumor cells.(Wolff AC, Hammond ME, Hicks DG, et al. Human Epidermal Growth Factor Receptor 2 Testing in Breast Cancer: American Society of Clinical Oncology/College of American Pathologists Clinical Practice Guideline Focused Update. J Clin Oncol. 2018;36(20):2105-2122 doi:10.1200/JCO.2018.77.8738)

PDF Report
Indicates whether the report includes an additional document with charts, images or other enriched information

No

Day(s) Performed
Outlines the days the test is performed. This field reflects the day that the sample must be in the testing laboratory to begin the testing process and includes any specimen preparation and processing time before the test is performed. Some tests are listed as continuously performed, which means that assays are performed multiple times during the day.

Monday through Friday

Report Available
The interval of time (receipt of sample at Mayo Clinic Laboratories to results available) taking into account standard setup days and weekends. The first day is the time that it typically takes for a result to be available. The last day is the time it might take, accounting for any necessary repeated testing.

5 to 15 days

Specimen Retention Time
Outlines the length of time after testing that a specimen is kept in the laboratory before it is discarded

Until 1 week after results are reported. Materials made at Mayo Clinic may be retained at Mayo Clinic indefinitely.

Performing Laboratory Location
Indicates the location of the laboratory that performs the test

Rochester

Fees
Several factors determine the fee charged to perform a test. Contact your U.S. or International Regional Manager for information about establishing a fee schedule or to learn more about resources to optimize test selection.

  • Authorized users can sign in to Test Prices for detailed fee information.
  • Clients without access to Test Prices can contact Customer Service 24 hours a day, seven days a week.
  • Prospective clients should contact their account representative. For assistance, contact Customer Service.

Test Classification
Provides information regarding the medical device classification for laboratory test kits and reagents. Tests may be classified as cleared or approved by the US Food and Drug Administration (FDA) and used per manufacturer instructions, or as products that do not undergo full FDA review and approval, and are then labeled as an Analyte Specific Reagent (ASR) product.

This test has been modified from the manufacturer's instructions. Its performance characteristics were determined by Mayo Clinic in a manner consistent with CLIA requirements. This test has not been cleared or approved by the US Food and Drug Administration.

CPT Code Information
Provides guidance in determining the appropriate Current Procedural Terminology (CPT) code(s) information for each test or profile. The listed CPT codes reflect Mayo Clinic Laboratories interpretation of CPT coding requirements. It is the responsibility of each laboratory to determine correct CPT codes to use for billing.

CPT codes are provided by the performing laboratory.

88360

LOINC® Information
Provides guidance in determining the Logical Observation Identifiers Names and Codes (LOINC) values for the order and results codes of this test. LOINC values are provided by the performing laboratory.

Test Id Test Order Name Order LOINC Value
HERMB HER Breast SemiQuant IHC + Reflex 85319-2
Result Id Test Result Name Result LOINC Value
Applies only to results expressed in units of measure originally reported by the performing laboratory. These values do not apply to results that are converted to other units of measure.
620801 Interpretation 50595-8
620802 Participated in the Interpretation No LOINC Needed
620803 Report electronically signed by 19139-5
620804 Material Received 81178-6
MA058 Fixed in 10% NB formalin w/in 1 hr 8100-0
MA059 Fixed in 10% NB formalin 6-72 hrs 8100-0
MA060 Tumor type 44638-5
MA061 Tumor classification 21918-8
MA062 Tissue was decalcified 8100-0
620805 Disclaimer 62364-5
620806 Case Number 80398-1

Test Setup Resources

Setup Files
Test setup information contains test file definition details to support order and result interfacing between Mayo Clinic Laboratories and your Laboratory Information System.

Excel | PHP Pdf | CMS Pdf

Sample Reports
Normal and Abnormal sample reports are provided as references for report appearance.

Normal Reports | Abnormal Reports

SI Sample Reports
International System (SI) of Unit reports are provided for a limited number of tests. These reports are intended for international account use and are only available through MayoLINK accounts that have been defined to receive them.

SI Normal Reports | SI Abnormal Reports